EDS00xx(3)		     One-Wire File System		    EDS00xx(3)

NAME
       EDS - Embedded Data Systems Sensors

       EDS0064
	      - Temperature

       EDS0065
	      - Temperature and Humidity

       EDS0066
	      - Temperature and Barometric Pressure

       EDS0067
	      - Temperature and Light

       EDS0068
	      - Temperature, Barometric Pressure and Light

       EDS0070
	      - Vibration

       EDS0071
	      - RTD 4-wire temperature

       EDS0072
	      - RTD 3-wire temperature

       EDS0080
	      - 8 Channel Analog Current Input

       EDS0082
	      - 8 Channel Analog Voltage Input

       EDS0083
	      - 4 Channel Analog Current Input

       EDS0085
	      - 4 Channel Analog Voltage Input

       EDS0090
	      - 8 Discrete I/O

SYNOPSIS
       Embedded Data Systems microprocessor-based sensors

   Sub-type ( all )
       7E [.]XXXXXXXXXXXX[XX][/[ device_id | device_type | tag ]]

   Memory ( all )
       7E [.]XXXXXXXXXXXX[XX][/[ memory | pages/page.[0-2|ALL] ]]

   Standard ( all )
       7E [.]XXXXXXXXXXXX[XX][/[ ]]

   Temperature ( EDS0064/5/6/7/8 )
       7E.XXXXXXXXXXXX/EDS006X	  [/[	temperature   |	  counters/seconds   |
       alarm/temp_[hi|low] | alarm/clear | set_alarm/temp_[hi|low]  |  thresh‐
       old/temp_[hi|low] ]]

   Humidity ( EDS0065/8 )
       7E.XXXXXXXXXXXX/EDS006X [/[ humidity | humidex | dew_point | heat_index
       |     alarm/humidity_[hi|low]	  |	 alarm/humidex_[hi|low]	     |
       alarm/dew_point_[hi|low]	 |  alarm/heat_index_[hi|low]  | alarm/clear |
       set_alarm/humidity_[hi|low]     |     set_alarm/humidex_[hi|low]	     |
       set_alarm/dew_point_[hi|low]  | set_alarm/heat_index_[hi|low] | thresh‐
       old/humidity_[hi|low] |
       threshold/humidex_[hi|low]  |  threshold/dew_point_[hi|low]  |  thresh‐
       old/heat_index_[hi|low] ]]

   Barometric Pressure ( EDS0066/8 )
       7E.XXXXXXXXXXXX/EDS006X [/[ pressure | inHg | alarm/pressure_[hi|low] |
       alarm/inHg_[hi|low]  |  alarm/clear  |  set_alarm/pressure_[hi|low]   |
       set_alarm/inHg_[hi|low]	 |   threshold/pressure_[hi|low]   |   thresh‐
       old/inHg_[hi|low] ]]

   Light ( EDS0067/8 )
       7E.XXXXXXXXXXXX/EDS006X [/[ light | alarm/light_[hi|low] |  alarm/clear
       | set_alarm/light_[hi|low] | threshold/light_[hi|low] ]]

   Vibration ( EDS0070 )
       7E.XXXXXXXXXXXX/EDS0070	[/[ vib_level | vib_peak | vib_min | vib_max |
       counter/seconds	  |	alarm/vib_[hi|low]     |     alarm/clear     |
       set_alarm/vib_[hi|low] | threshold/vib_[hi|low] ]]

   RTD Temperature ( EDS0071/2 )
       7E.XXXXXXXXXXXX/EDS007X	[/[  temperature  | resistance | raw | delay |
       user_byte | calibration/[key|constant]  |  counter/[seconds|samples]  |
       alarm/temp_[hi|low]    |	   alarm/RTD_[hi_low]	 |    alarm/clear    |
       set_alarm/temp_[hi|low]	  |    set_alarm/RTD_[hi_low]	 |     thresh‐
       old/temp_[hi|low] | threshold/RTD_[hi_low] ]]

   Analog Current ( EDS0080 [0-7] EDS0083 [0-3] )
       7E.XXXXXXXXXXXX/EDS0080 [/[ current.[0-7|ALL] | max_current.[0-7|ALL] |
       min_current.[0-7|ALL] | threshold/current_hi.[0-7|ALL] | threshold/cur‐
       rent_low.[0-7|ALL]   |	alarm/current_hi.[0-7|ALL|BYTE]	 |  alarm/cur‐
       rent_low.[0-7|ALL|BYTE]	 |    set_alarm/current_hi.[0-7|ALL|BYTE]    |
       set_alarm/current_low.[0-7|ALL|BYTE]   |	 counter/seconds  |  memory  |
       pages/page.[0-4|ALL]

   Analog Voltage ( EDS0082 [0-7] EDS0085 [0-3] )
       7E.XXXXXXXXXXXX/EDS0082 [/[  volts.[0-7|ALL]  |	max_volts.[0-7|ALL]  |
       min_volts.[0-7|ALL]    |	   threshold/volts_hi.[0-7|ALL]	   |   thresh‐
       old/volts_low.[0-7|ALL]	   |	  alarm/volts_hi.[0-7|ALL|BYTE]	     |
       alarm/volts_low.[0-7|ALL|BYTE]  |  set_alarm/volts_hi.[0-7|ALL|BYTE]  |
       set_alarm/volts_low.[0-7|ALL|BYTE]  |  counter/seconds	|   memory   |
       pages/page.[0-4|ALL]

   I/O ( EDS0090 )
       7E.XXXXXXXXXXXX/EDS009X	 [/[   input.[0-7|ALL]	|  counter/[seconds  |
       pulses.[0-7|ALL]|reset.[0-7|ALL|BYTE]]		   |		  out‐
       put/[set.[0-7|ALL|BYTE]|off.[0-7|ALL|BYTE]|reset.[0-7|ALL|BYTE]]	     |
       latch/[state.[0-7|ALL|BYTE]|reset.[0-7|ALL|BYTE]]		     |
       alarm/[hi.[0-7|ALL]|low.[0-7|ALL]|clear]				     |
       set_alarm/[hi.[0-7|ALL]|low.[0-7|ALL]]

   LED ( all )
       7E.XXXXXXXXXXXX/EDS00XX/LED/[state|control]

   relay ( all )
       7E.XXXXXXXXXXXX/EDS00XX/relay/[state|control]

FAMILY CODE
       7E

TEMPERATURE
   EDS006X/temperature
       read-only, floating-point
       Temperature for the EDS006X series of chips. In the -55C to 125C range,
       with  0.0636C precision. The temperature is read every second, continu‐
       ally. Data is presented in the specified temperature scale, Celsius  by
       default.

   EDS007X/temperature
       read-only, floating-point
       Temperature  for	 the  EDS007X series of chips. Read using a wide-range
       precise RTD sensor. Typical range is -60C to 260C with 0.15C  accuracy,
       although	 a range of -200C to 850C is possible. The temperature is read
       every second unless a longer EDS007X/delay is given. Data is  presented
       in the specified temperature scale, Celsius by default.

   EDS007X/resistance
       read-only, floating-point
       Actual measured resistance (Ohms) of the RTD element. Useful if the RTD
       element doesn't conform to the typical European IEC 60751 standard.

HUMIDITY
   EDS006X/humidity
       read-only, floating-point
       Relative humidity in the 0-100 range (percent). Read every 0.2 seconds.

   EDS006X/dew_point
       read-only, floating-point
       Dew point computed from EDS006X/temperature and	EDS006X/humidity  com‐
       putered	every  0.2  seconds.  Data  is a calculated temperature and is
       reported in the specified temperature scale. Default Celsius.

   EDS006X/heat_index
       read-only, floating-point
       Heat index computed from EDS006X/temperature and EDS006X/humidity  com‐
       putered	every  0.2  seconds.  Data  is a calculated temperature and is
       reported in the specified temperature scale. Default Celsius.

   EDS006X/humidex
       read-only, floating-point
       Humidex (popular	 in  Canada)  computed	from  EDS006X/temperature  and
       EDS006X/humidity	 computed  every  0.2  seconds.	 Data is a percent and
       reported in the 0-100 range.

LIGHT
   EDS006X/light
       read-only, unsigned integer
       Ambient light in Lux. Measured every 0.2 seconds.

PRESSURE
   EDS006X/pressure
       read-only, floating-point
       Ambient pressure, measured every 0.2  seconds.  Data  in	 the  selected
       pressure scale (default mBar).

   EDS006X/inHg
       read-only, floating-point
       Ambient pressure, measured every 0.2 seconds. Data in the inHg scale.

VIBRATION
   EDS0070/vib_level
       read_only, unsigned integer
       Vibration registered by sensor (instantaneous value) in 0-1023 range.

   EDS0070/vib_peak
       read_only, unsigned integer
       Vibration  registered by sensor (highest recent value -- slowly decays)
       in 0-1023 range.

   EDS0070/vib_min
       read_only, unsigned integer
       Vibration registered by sensor (lowest value) in 0-1023 range.

   EDS0070/vib_max
       read_only, unsigned integer
       Vibration registered by sensor (highest value) in 0-1023 range.

CURRENT
       The EDS0080 and EDS0083 measure current in the 4-20mA range, with upper
       and  threshold limits and the ability to set an alarm and independently
       trigger a relay if the value is out of range.

   EDS0080/current.[0-7|ALL] EDS0083/current.[0-3|ALL]
       read-only, floating point
       Current current level. (4-20mA)

   EDS0080/max_current.[0-7|ALL] EDS0080/max_current.[0-3|ALL]
   EDS0083/min_current.[0-7|ALL] EDS0083/min_current.[0-3|ALL]
       read-only, floating point
       Maximum and minimum current readings since last alarm/clear command.

VOLTAGE
       The EDS0082 and EDS0085 measure voltage in the 0-10V range, with	 upper
       and  threshold limits and the ability to set an alarm and independently
       trigger a relay if the value is out of range.

   EDS0082/volts.[0-7|ALL] EDS0085/volts.[0-3|ALL]
       read-only, floating point
       Current voltage level. (0-10V)

   EDS0082/max_volts.[0-7|ALL] EDS0085/max_volts.[0-3|ALL]
   EDS0082/min_volts.[0-7|ALL] EDS0085/min_volts.[0-3|ALL]
       read-only, floating point
       Maximum and minimum voltage readings since last alarm/clear command.

THRESHOLD
       High and low range of acceptable sensor readings. Values	 outside  this
       range will trigger an alarm if the corresponding EDS00XX/set_alarm flag
       is set.

   EDS00XX/threshold/temp_hi EDS00XX/threshold/temp_low
       read-write, floating-point
       Threshold temperatures in the specified temperature scale. Default Cel‐
       sius.

   EDS006X/threshold/humidity_hi EDS006X/threshold/humidity_low
   EDS006X/threshold/dew_point_hi EDS006X/threshold/dew_point_low
   EDS006X/threshold/heat_index_hi EDS006X/threshold/heat_index_low
   EDS006X/threshold/humidex_hi tEDS006X/threshold/humidex_low
       read-write, floating-point
       Threshold  humidity values.  Dew point and Heat index are in the speci‐
       fied temperature scale. Default Celsius.

   EDS006X/threshold/pressure_hi EDS006X/threshold/pressure_low
   EDS006X/threshold/inHg_hi EDS006X/threshold/inHg_low
       read-write, floating-point
       Threshold barometric pressure values.  Pressure is the specified	 pres‐
       sure scale. Default mBar.

   EDS006X/threshold/light_hi EDS006X/threshold/light_low
       read-write, unsigned
       Threshold light (lux) values.

   EDS0070/threshold/vib_hi EDS0070/threshold/vib_low
       read-write, unsigned
       Vibration sensor alarm limits in 0-1023 range.

   EDS007X/threshold/resistance_hi EDS007X/threshold/resistance_low
       read-write, floating-point
       Threshold RTD resistance values (Ohm).

   EDS0080/threshold/current_hi.[0-7|ALL]		EDS0080/threshold/cur‐
       rent_low.[0-7|ALL]
   EDS0082/threshold/volts_hi.[0-7|ALL] EDS0082/threshold/volts_low.[0-7|ALL]
   EDS0083/threshold/current_hi.[0-3|ALL		EDS0083/threshold/cur‐
       rent_low.[0-3|ALL]
   EDS0085/threshold/volts_hi.[0-3|ALL] EDS0085/threshold/volts_low.[0-3|ALL|]
       read-write, floating-point
       Voltage or current threshold limits

SET ALARM
       Set  conditional	 alarm	to  trigger if corresponding flag is set. Also
       must set high and low threshold.

   EDS00XX/set_alarm/temp_hi EDS00XX/set_alarm/temp_low
   EDS006X/set_alarm/humidity_hi EDS006X/set_alarm/humidity_low
   EDS006X/set_alarm/dew_point_hi EDS006X/set_alarm/dew_point_low
   EDS006X/set_alarm/heat_index_hi EDS006X/set_alarm/heat_index_low
   EDS006X/set_alarm/humidex_hi EDS006X/set_alarm/humidex_low
   EDS006X/set_alarm/pressure_hi EDS006X/set_alarm/pressure_low
   EDS006X/set_alarm/inHg_hi EDS006X/set_alarm/inHg_low
   EDS006X/set_alarm/light_hi EDS006X/set_alarm/light_low
   EDS0070/set_alarm/vib_hi EDS0070/set_alarm/vib_low
   EDS007X/set_alarm/resistance_hi EDS007X/set_alarm/resistance_low
   EDS0080/set_alarm/current_hi.[0-7|ALL|BYTE]		EDS0080/set_alarm/cur‐
       rent_low.[0-7|ALL|BYTE]
   EDS0082/set_alarm/volts_hi.[0-7|ALL|BYTE]
       EDS0082/set_alarm/volts_low.[0-7|ALL|BYTE]
   EDS0083/set_alarm/current_hi.[0-3|ALL|BYTE]		EDS0083/set_alarm/cur‐
       rent_low.[0-3|ALL|BYTE]
   EDS0085/set_alarm/volts_hi.[0-3|ALL|BYTE]
       EDS0085/set_alarm/volts_low.[0-3|ALL|BYTE]
       read-write, yes-no
       Flag to set corresponding out-of-range alarm. "1" turns	on  flag,  and
       "0" turns off.

ALARM
       Show or clear the out-of-range flags. When set, the device will respond
       to a conditional search. Clearing  the  flags  is  performed  with  the
       EDS00XX/alarm/clear  command, not just having the measured value return
       to the center range.

   EDS00XX/alarm/clear
       write-only, yes-no
       Write "1" to clear ALL the alarm flags.

   EDS00XX/alarm/temp_hi EDS00XX/alarm/temp_low
   EDS006X/alarm/humidity_hi EDS006X/alarm/humidity_low
   EDS006X/alarm/dew_point_hi EDS006X/alarm/dew_point_low
   EDS006X/alarm/heat_index_hi EDS006X/alarm/heat_index_low
   EDS006X/alarm/humidex_hi EDS006X/alarm/humidex_low
   EDS006X/alarm/pressure_hi EDS006X/alarm/pressure_low
   EDS006X/alarm/inHg_hi EDS006X/alarm/inHg_low
   EDS006X/alarm/light_hi EDS006X/alarm/light_low
   EDS0070/alarm/vib_hi EDS007X/alarm/vib_low
   EDS007X/alarm/resistance_hi EDS007X/alarm/resistance_low
   EDS0080/alarm/current_hi.[0-7|ALL|BYTE]		    EDS0080/alarm/cur‐
       rent_low.[0-7|ALL|BYTE]
   EDS0082/alarm/volts_hi.[0-7|ALL|BYTE]
       EDS0082/alarm/volts_low.[0-7|ALL|BYTE]
   EDS0083/alarm/current_hi.[0-3|ALL|BYTE]		    EDS0083/alarm/cur‐
       rent_low.[0-3|ALL|BYTE]
   EDS0085/alarm/volts_hi.[0-3|ALL|BYTE]
       EDS0085/alarm/volts_low.[0-3|ALL|BYTE]
   EDS0090/alarm/hi[0-7|ALL] EDS0090/alarm/low[0-7|ALL]
       read-only, yes-no
       Show corresponding out-of-range alarm. "1" means in  alarm  state.  Can
       only be turned off with clear

RELAY
       Optionally found on the EDS006X and EDS007X. Can be controlled by soft‐
       ware or the alarm condition.

   EDS00XX/relay/control
       read-write, unsigned
       Set the relay control scheme:

       0      alarm control with hysteresis

       1      alarm control but need EDS00XX/alarm/clear to unset

       2      Control with EDS00XX/relay/state

       3      Always off

   EDS00XX/relay/state
       read-write, yes-no
       Turn the relay on or off if the EDS00XX/relay/control is set to 2

LED light
       Found on the EDS006X and EDS007X. Can be controlled by software or  the
       alarm condition.

   EDS00XX/LED/control
       read-write, unsigned
       Set the LED control scheme:

       0      alarm control with hysteresis

       1      alarm control but need EDS00XX/alarm/clear to unset

       2      Control with EDS00XX/LED/state

       3      Always off

   EDS00XX/LED/state
       read-write, yes-no
       Turn the LED on or off if the EDS00XX/LED/control is set to 2

COUNTER
   EDS00XX/counter/seconds
       read-only, unsigned integer
       Approximate seconds since power up.

   EDS0090/counter/pulses.0-7 | ALL
       read-only, unsigned integer
       Pulses  on  each channel since power up or reset. Channel should be set
       to input ( EDS0090/output/set.x =0 )

   EDS0090/counter/reset.0-7 | ALL | BYTE
       read-write, yes-no
       Clear the pulse count on the corresponding channel.

   EDS007X/counter/samples
       read-only, unsigned integer
       Approximate samples since power up.

MEMORY
   memory
       read-write, binary
       96 bytes of memory. Not all is writable. Access to EDS00XX functions is
       better  accomplished through the data fields, which expose all the chip
       features.

   pages/page.0 ... pages/page.2 pages/page.ALL
       read-write, binary
       Memory is split into 3 pages of 32 bytes each. Only page 2 is writable.
       ALL is an aggregate of the pages. Each page is accessed sequentially.

   EDS008X/memory
       read-write, binary
       160  bytes  of memory. Not all is writable. Access to EDS008X functions
       is better accomplished through the data fields, which  expose  all  the
       chip  features.	The EDS008X has more memory than the other chips, so a
       separate entry in the sudbirectory makes it available.

   EDS008X/pages/page.0-4 EDS008X/pages/page.ALL
       read-write, binary
       Memory is split into 5 pages of 32 bytes each. Only page 5 is writable.
       The  EDS008X  has more memory than the other chips, so a separate entry
       in the sudbirectory makes the extra pages available.  ALL is an	aggre‐
       gate of the pages. Each page is accessed sequentially.

SUB-TYPE
   tag
       read-only, ascii
       Text description of the device. Up to 30 characters.

   device_type
       read-only, ascii
       Device name. E.g: "EDS0071"

   device_id
       read-only, unsigned
       Number corresponding to the hexidecimal type. E.g:
       EDS0071 -> 0x0071 -> 113 in decimal

STANDARD PROPERTIES
ALARMS
       None.

DESCRIPTION
   EDS00XX Overview
       The EDS00XX sensors are a family of devices from Embedded Data Systems.
       Although provisional support for several types is included, only	 early
       EDS0071 has been tested with OWFS.

       The sensors all share the same 7E family code. They have the same first
       memory page layout with a text tag property that describes them, and  a
       sub-type	 field giving the particular sensor. The data properties shown
       in a directory listing will be specific to the sensor sub-type.

   EDS0071
       The EDS0071 (3) is a micro-processor based 1-wire  slave	 made  by  EDS
       (Embedded Data Systams). Is uses a high accuracy and extended range RTD
       sensor to accurately measure temperatures.

   Memory
       The EDS00XX sensors are read and controlled by accessing parts of their
       on-chip	memory.	 Although  the	raw memory contents are exposed in the
       memory and pages	 fields,  all  the  available  functionality  is  best
       accessed	 by  the  other fields. In fact, directly changing memory will
       confuse OWFS.

   Relay and LED
       All the EDS00XX sensors have an LED and optional	 relay.	 Both  can  be
       controlled either directly, or when the sensors reaches an alarm state.

ADDRESSING
DATASHEET
       EDS0064 EDS0065 EDS0066 EDS0067 EDS0068
	      https://docs.google.com/viewer?a=v&pid=explorer&chrome=true&srcid=0B_K3iXL‐
	      LAI7vN2E4NGUwMGItMzc5MS00NDNmLTljZGYtN2NmM2Q3YWE5NmVh&hl=en

       EDS0070
	      http://www.embeddeddatasystems.com/assets/images/support‐
	      Files/manuals/EN-UserMan%20%20OW-Vibration%20V01.pdf

       EDS0071 EDS0072
	      https://docs.google.com/viewer?a=v&pid=explorer&chrome=true&srcid=0B_K3iXL‐
	      LAI7vMjVhNWQzYzItZTRhMi00NGY2LWI4NTQtYjZlYTgzYzY0NmNm&hl=en

       EDS0080 EDS0083
	      https://eds.zendesk.com/attach‐
	      ments/token/ulnpj6mvpp6yzwo/?name=EN-USERMAN+OW-IO-
	      AIX-420+v1p0.pdf

       EDS0082 EDS0085
	      https://eds.zendesk.com/attach‐
	      ments/token/vyai6ylpw6edala/?name=EN-USERMAN+OW-IO-
	      AIX-10V+v1p0.pdf

       EDS0090
	      8 Discrete I/O

SEE ALSO
AVAILABILITY
       http://www.owfs.org

AUTHOR
       Paul Alfille (paul.alfille@gmail.com)

OWFS Manpage			     2003			    EDS00xx(3)
[top]

List of man pages available for DragonFly

Copyright (c) for man pages and the logo by the respective OS vendor.

For those who want to learn more, the polarhome community provides shell access and support.

[legal] [privacy] [GNU] [policy] [cookies] [netiquette] [sponsors] [FAQ]
Tweet
Polarhome, production since 1999.
Member of Polarhome portal.
Based on Fawad Halim's script.
....................................................................
Vote for polarhome
Free Shell Accounts :: the biggest list on the net